| SPECIFICATIONS |
| Measurement Ranges |
| Technique: |
Advanced Dual Sensor Technology (ADST) |
| CO Ranges: |
0 - 10 ppm up to 0 - 40 000 ppm (4 %) |
| NO Ranges: |
0 - 10 ppm up to 0 - 50 000 ppm (5 %) |
| NO2 Ranges: |
0 - 10 ppm up to 0 - 1 000 ppm (0.1 %) |
| SO2 Ranges: |
0 - 10 ppm up to 0 - 10 000 ppm (1 %) |
| CO2 Ranges: |
0 - 3 Vol. % up to 0 - 100 Vol. % |
| Resolution: |
0.1 ppm (CO, NO, NO2 & SO2); 0.1 % (CO2) |
| Linearity: |
< 2 % of range |
| Zero drift: |
< 2 % of range per month |
| Span drift: |
< 2 % of range per month |
| O2 Ranges: |
0 - 5 Vol % to 0 - 25 Vol % |
| Resolution: |
0.01 Vol % |
| Linearity: |
< 0.2 Vol % |
| Zero drift: |
< 0.2 Vol % per month |
| Span drift: |
< 0.2 Vol % per month |
| Calibration |
| Calibration method: |
Manual or Automatic
2 or 3 point span calibration; zero calibration
Automatic Gas Selection (option)
Microprocessor controlled |
| Display: |
|
| Type: |
LCD + LED backlight |
| Size: |
125 x 35 mm / 5 x 1.5 in |
| Parameters: |
5 function keys; alphanumeric and graphic |
| Outputs/Inputs |
| Analogue outputs: |
Isolated current loops
One per gas (plus total NOx if applicable)
expandable up to 32, in multiples of 8
0,2 or 4 to 10 or 20 mA |
| Relays outputs: |
One per gas (plus total NOx if applicable) plus System OK, Warning, Calibrating, Maintenance, expandable up to 20, in multiples of 5, including calibration gas status relays |
| Relay rating: |
Isolated change-over contacts rated 0.5 A @ 125 V a.c.;
1 A @ 30 V d.c. |
| Digital Inputs: |
Calibrate, Standby, Optional fuel selection
Volt-free contact closure or 5 - 24 V d.c. |
| Analogue inputs: |
Isolated current loops
Up to 16, in multiples of 8, active or passive
0, 2 or 4 to 10 or 20 mA |
| Communications: |
RS485 Modbus serial digital data bus ( 2 or 4 wire) |
| Environmental |
| Enclosure: |
Stainless steel sealed to the requirements of
IP65 / NEMA 4X |
Operating (ambient)
temperature: |
-40 &de°C / -40 °F to 104 °F standard
-40 &de°C / -40 °F to 122 °F most measurement ranges |
| Compliance |
Measurement
standards: |
Meets the requirements of ISO 12039, ISO 10849 & ISO 7935
Meets LCP & Waste Incineration Directives 2001/80/EC and 2000/76/EC
Meets EN14181 QAL 1 requirements for most processes
Meets the requirements for USEPA Part 60, 63, 75, 266, 267 & 503 applications |
| Approvals: |
TÜV approved to the 13th, 17th & 27th BImSchV and TA Lüft
MCerts certified to Performance Standards for Continuous Emissions Monitoring Systems (Nov.1998) |
| Electrical Safety: |
Conforms to EN-61010-2 |
| EMC: |
Conforms to EN61326 (Industrial) |
| Power |
| Power Supply: |
100 to 120 V a.c. or 200 to 240 V a.c., 50 - 60 Hz |
| Power Consumption: |
Analyser: 600 W
Anti-freeze sample line: 15 W / m (5 W / ft) |
| Note: Additional power and air required for ChillerProbeTM |
| Gas & Air Requirements |
Instrument air
For zero calibration: |
4 bar / 60 psi min clean,dry and solvent free
1 bar / 15 psi; 10 l/min / 0.4 cfm maximum |
Calibration gas
(recommended): |
1 bar / 15 psi; 2 l/min / 0.08 cfm
10 litres / 0.35 cu. ft. (16 litres / 0.56 cu. ft. for NO2) per calibration approx. |
| Calibration gas type: |
Specific to each gas type and measurement span |
Dimensions
(H x W x D) |
Single stream: 600 x 600 x 420 mm / 24 x 24 x 17in
Twin stream: 870 x 600 x 420 mm / 34 x 24 x 17 in |
| Weight: |
Single stream: 50 kg /110 lb
Twin stream: 73 kg / 161 lb |
| Options |
| Automatic Calibration |
| Twin Stream Option |
| Remote Diagnostics |
Continuous Product Development may make it necessary
to change these details without notice
|