| specifications |
| Sample Conditioning |
Filtration:
|
Heated unit with replaceable element
180 °C / 356 °F suitable for most applications
280 °C / 536°F for applications where ammonia is present in the flue gas |
| Filter Element: |
1.0 µm ceramic |
| Sample Cooler: |
2-stage cooler system (incorporating Peltier unit) with automatic condensate removal |
| Blow-back |
| Method:
|
Heated, filtered, compressed air |
| Triggers |
Internal timer; external trigger (via digital link); Manual (push button); +24 V external drive |
| Temperature: |
150 °C / 300 °F |
| Timer Interval: |
15 minutes to 24 hours user selectable |
| Duration: |
3 sec's |
| Control System |
| Communications: |
Remote control via RS 485 digital link |
| Automatic Controls: |
Filter temperature; Cooler temperature; Condensate removal; Blow-back timing; Blow-back air temperature |
| Manual Controls: |
Blow-back trigger |
| Status LEDs: |
Power; Blow-back; System OK; Communications
|
Status Outputs (via
digital link): |
Filter temperature; Blow-back air temperature; Blow-back pressure; Cooler Temperature; Enclosure temperature; Condensate pumps active; Sample gas temperature (optional) |
Alarms (indicated
at analyzer): |
Filter temperature (low/high)
Blow-back accumulator temperature (low/high)
Cooler temperature (low/high)
Blow-back accumulator low pressure
Moisture slip
Pump fault |
| Relay Outputs: |
System OK; Blow-back in progress; Filter cold; Cooler fault |
| Additional Output: |
Power for anti-freeze condensate sample line |
| Environmental |
| Enclosure: |
Stainless steel sealed to the requirements of IP 65 / NEMA 4X |
Operating
Temperature: |
-25 &de°C / -13 to +131 °F (option to -40 °C / -40 °F) |
| Compliance |
| EMC: |
Conforms to EN-61326 (Class A) |
| Electrical Safety: |
Conforms to EN-61010-2 |
| Power |
| Power Supply: |
110 or 230 V a.c., ±10%, 50 or 60 Hz |
| Power Consumption: |
900 W |
| Air Requirements |
Air Requirements
for cooling: |
Dependent upon ambient temperature and process conditions |
Dimensions
(H x W x D) |
685 x 510 x 356 mm /
27 x 20 x 14 in |
| Weight: |
56 kg / 124 lb |
| Mounting: |
Single flange - 3" ANSI, 150 lb type |
| Flange Temperature: |
up to 180 °C / 356 °F without adaptor
up to 500 °C / 932 °F with adaptor |
Optional Flange
Adaptor: |
2", 3", or 4" ANSI 150 lb; DN65 PN6 |
| Probe Pipes and Materials |
316 Stainless Steel
Hastelloy C-276
Hastelloy X
Mullite |
up to 600 °C / 1112 °F
up to 900 °C / 1652 °F
up to 1200 °C / 2200 °F
up to 1700 °C / 3100 °F |
| Lengths*: |
0.5 m / 1.6 ft; 1.0 m / 3.3 ft; 1.5 m / 5 ft; 2.0 m / 6.6 ft |
Probe Pre-filter#:
|
2.0 µm Sintered stainless steel (Temp. <400 °C / 750 °F)
2.0 µm Hastelloy (Temp. up to 800 °C / 1470 °F) |
* Mullite probes only available in 0.5 and 1.0m lengths
# Available with all probe types except mullite |
| Options |
Sample Gas Cooler
Temp. sensor:
Air blower for
instrument cooling: |
Type K thermocouple
Low capacity or high capacity blower units |
| |
|
| Continuous Product Development may make it necessary to change these details without notice |