Specification |
|
| |
Type: |
Land Precision Glass Thermometer Type JRG/A, mounted in a water cooled probe Type P/J. |
|
| |
Detector: |
Thermopile |
|
| |
Optical components: |
Infrared filter to give an effective waveband of 4.8 to 5.2µm. Zinc sulphide lens and
fluorite window. |
|
| |
Lens aperture: |
12.5mm/1/2in. |
|
| |
Sighting: |
The thermometer is focused at 75mm/3in from the front of the lens and the nominal target
size at this distance is 25 mm/1 in diameter. |
|
| |
Output: |
Amplified voltage signal 6000mV at 1200°C black body. Linear in radiance. |
|
| |
Range: |
200 to 1200°C/300 to 1100°C |
|
| |
Response time: |
1s to 98% of an instantaneous change after setting time of 15s on first being introduced to
hot surroundings. |
|
| |
Ambient operating temperature: |
0 to 70°C/32 to 158°F (inside a furnace) |
|
| |
Absolute accuracy: |
0.75%K + 1°C |
|
| |
Repeatability: |
1°C |
|
| |
Ancillary services: |
Water - flow required depends on the length of probe and temperature of hot zone.
Air - a flow of filtered low pressure air equivalent to 1 litre/min/2ft3/h. |
|
| |
Connections: |
Electrical - self locating demountable connector.
Water - snap on connectors.
Air - snap on connectors. |
|
| |
Dimensions: |
43mm/1.7in x 28mm/1.1in dia. |
|
| |
Weight (Probe and JRG/A): |
Approx 4.5kg/10lb plus 1.8 kg/4lb per 300mm/1ft of operating length. |
|